{| class="wikitable"
! colspan="2" |
== '''Overview''' ==
|-
|'''Department'''
|Point of Care Testing (POCT)
|-
|'''Test Code'''
|N/A
|-
|'''Synonyms'''
|HbA1c whole blood (POCT)
glycated haemoglobin
HBA
|-
|'''Availability'''
|Cobas b101 HbA1c devices are available at both Taranaki Base and Hawera hospital.
|-
|'''Turnaround Time'''
|6 minutes from placing cartridge into analyser with patients' blood sample
|-
|'''Stability – Add test'''
|60 seconds once blood sample has been applied to cartridge disc.
EDTA samples within 2 hours from collection
Lithium-Heparin within 8 hours from collection
|-
! colspan="2" |
== '''Specimen Requirements''' ==
|-
|'''Specimen'''
|Whole Blood  
|-
|'''Preferred'''
|Finger Stab - Droplet of Whole blood
|-
|'''Acceptable'''
|Purple top (EDTA) or Lithium-Heparin, non-gel (green top)
|-
|'''Unacceptable'''
|Samples containing anticoagulants or other additives
Separated samples
|-
|'''Minimum Volume'''
|2µL whole blood
|-
|'''Specimen Collection and Transportation'''
|Clean and dry patients' hand and wipe away the first drop of blood.
|-
! colspan="2" |
== '''Test Information''' ==
|-
|'''Test Method'''
|Cobas b101 - ROCHE analyser
|-
|'''Unit of Measurement'''
| mmol/mol
|-
|'''Reference Intervals'''
|National diagnostic threshold for HbA1c (as of July 1st 2026)
- '''Normal:''' <42 mmol/mol
- '''Prediabetes:''' 42-47 mmol/mol
- '''Diabetes''': ≥48 mmol/mol
|-
|'''Uncertainty of measurement'''
|N/A
|-
|'''Additional Information'''
|No confirmatory test now required if ≥53 mmol/mol
Confirmatory test required as soon as practical if HbA1c 48-52 mmol/mol eg repeat HbA1c, fasting glucose or random glucose (if symptomatic)
Ranges of measurement 20-130 mmol/mol
The presence of some haemoglobin variants may require testing via alternative methods for HbA1c or the laboratory may refer clinician to glucose based criteria for diagnosis and monitoring.
|-
! colspan="2" |
== '''Comments''' ==
|-
| colspan="2" |Quick user guides for testing should be available with or near the analysers.
All procedures, worksheets and competency checklist for sign-off are available on the intranet to all Health NZ Staff.
This test is covered under the current IANZ Laboratory accreditation.
|}