Tests:HbA1c (POCT): Difference between revisions
Tests:HbA1c (POCT)
No edit summary |
m Mark as reviewed |
||
| (11 intermediate revisions by 3 users not shown) | |||
| Line 2: | Line 2: | ||
[[Category:Tests]] | [[Category:Tests]] | ||
[[Category:POCT]] | [[Category:POCT]] | ||
[[Category:Hawera]] | |||
{| class="wikitable" | |||
! colspan="2" | | |||
== '''Overview''' == | |||
|- | |||
|'''Department''' | |||
|Point of Care Testing (POCT) | |||
|- | |||
|'''Test Code''' | |||
|N/A | |||
|- | |||
|'''Synonyms''' | |||
|HbA1c whole blood (POCT) | |||
glycated haemoglobin | |||
HBA | |||
|- | |||
|'''Availability''' | |||
|Cobas b101 HbA1c devices are available at both Taranaki Base and Hawera hospital. | |||
|- | |||
|'''Turnaround Time''' | |||
|6 minutes from placing cartridge into analyser with patients' blood sample | |||
|- | |||
|'''Stability – Add test''' | |||
|60 seconds once blood sample has been applied to cartridge disc. | |||
EDTA samples within 2 hours from collection | |||
Lithium-Heparin within 8 hours from collection | |||
|- | |||
! colspan="2" | | |||
== '''Specimen Requirements''' == | |||
|- | |||
|'''Specimen''' | |||
|Whole Blood | |||
|- | |||
|'''Preferred''' | |||
|Finger Stab - Droplet of Whole blood | |||
|- | |||
|'''Acceptable''' | |||
|Purple top (EDTA) or Lithium-Heparin, non-gel (green top) | |||
|- | |||
|'''Unacceptable''' | |||
|Samples containing anticoagulants or other additives | |||
Separated samples | |||
|- | |||
|'''Minimum Volume''' | |||
|2µL whole blood | |||
|- | |||
|'''Specimen Collection and Transportation''' | |||
|Clean and dry patients' hand and wipe away the first drop of blood. | |||
|- | |||
! colspan="2" | | |||
== '''Test Information''' == | |||
|- | |||
|'''Test Method''' | |||
|Cobas b101 - ROCHE analyser | |||
|- | |||
|'''Unit of Measurement''' | |||
| mmol/mol | |||
|- | |||
|'''Reference Intervals''' | |||
|National diagnostic threshold for HbA1c (as of July 1st 2026) | |||
- '''Normal:''' <42 mmol/mol | |||
- '''Prediabetes:''' 42-47 mmol/mol | |||
- '''Diabetes''': ≥48 mmol/mol | |||
|- | |||
|'''Uncertainty of measurement''' | |||
|N/A | |||
|- | |||
|'''Additional Information''' | |||
|No confirmatory test now required if ≥53 mmol/mol | |||
Confirmatory test required as soon as practical if HbA1c 48-52 mmol/mol eg repeat HbA1c, fasting glucose or random glucose (if symptomatic) | |||
Ranges of measurement 20-130 mmol/mol | |||
The presence of some haemoglobin variants may require testing via alternative methods for HbA1c or the laboratory may refer clinician to glucose based criteria for diagnosis and monitoring. | |||
|- | |||
! colspan="2" | | |||
== '''Comments''' == | |||
|- | |||
| colspan="2" |Quick user guides for testing should be available with or near the analysers. | |||
All procedures, worksheets and competency checklist for sign-off are available on the intranet to all Health NZ Staff. | |||
This test is covered under the current IANZ Laboratory accreditation. | |||
|} | |||