Tests:HbA1c (POCT): Difference between revisions
Tests:HbA1c (POCT)
No edit summary |
m Mark as reviewed |
||
| (6 intermediate revisions by 2 users not shown) | |||
| Line 2: | Line 2: | ||
[[Category:Tests]] | [[Category:Tests]] | ||
[[Category:POCT]] | [[Category:POCT]] | ||
[[Category:Hawera]] | |||
{| class="wikitable" | {| class="wikitable" | ||
! colspan="2" | | ! colspan="2" | | ||
| Line 60: | Line 61: | ||
|- | |- | ||
|'''Reference Intervals''' | |'''Reference Intervals''' | ||
| | |National diagnostic threshold for HbA1c (as of July 1st 2026) | ||
- '''Normal:''' <42 mmol/mol | |||
- '''Prediabetes:''' 42-47 mmol/mol | |||
- '''Diabetes''': ≥48 mmol/mol | |||
|- | |- | ||
|'''Uncertainty of measurement''' | |'''Uncertainty of measurement''' | ||
| Line 71: | Line 72: | ||
|- | |- | ||
|'''Additional Information''' | |'''Additional Information''' | ||
|Ranges of measurement 20- | |No confirmatory test now required if ≥53 mmol/mol | ||
Confirmatory test required as soon as practical if HbA1c 48-52 mmol/mol eg repeat HbA1c, fasting glucose or random glucose (if symptomatic) | |||
Ranges of measurement 20-130 mmol/mol | |||
The presence of some haemoglobin variants may require testing via alternative methods for HbA1c or the laboratory may refer clinician to glucose based criteria for diagnosis and monitoring. | The presence of some haemoglobin variants may require testing via alternative methods for HbA1c or the laboratory may refer clinician to glucose based criteria for diagnosis and monitoring. | ||
|- | |- | ||
| Line 86: | Line 91: | ||
| review_ping = | | review_ping = | ||
}} | }} | ||
<!-- review-ping: 2026-09-01T21:13:58.707Z --> | |||
<!-- review-ping: 2026-09-01T21:16:20.808Z --> | |||