Tests:Digoxin, Plasma: Difference between revisions
Tests:Digoxin, Plasma
No edit summary |
Updated method Alinity |
||
| (2 intermediate revisions by 2 users not shown) | |||
| Line 72: | Line 72: | ||
|- | |- | ||
|Test Method | |Test Method | ||
|Abbott Alinity | |Abbott Alinity c series analyser using Abbott reagents. Particle-enhanced turbidimetric inhibition immunoassay (PETINIA) | ||
|- | |- | ||
|Advanced Notice | |Advanced Notice | ||
| Line 83: | Line 83: | ||
== Comments == | == Comments == | ||
|- | |- | ||
| colspan="2" |Specimens from patients treated with Digoxin immune FAB (Digibind) should be viewed with caution, as this substance is known to inhibit the assay. | | colspan="2" |Specimens from patients treated with Digoxin immune FAB (Digibind) should be viewed with caution, as this substance is known to inhibit the assay causing false elevated values. | ||
Heterophilic antibodies have been shown to interact with immunoglobulin reagents in some immunoassay systems. | Heterophilic antibodies have been shown to interact with immunoglobulin reagents in some immunoassay systems. | ||
|} | |} | ||
{{DocMeta | |||
| owner = Crossley1 | |||
| review_ping = | |||
}} | |||