HbA1c (POCT)

Revision as of 14:32, 1 October 2025 by Pam Chung (talk | contribs)

Tests:HbA1c (POCT)

Overview

Department Point of Care Testing (POCT)
Test Code N/A
Synonyms HbA1c whole blood (POCT)

glycated haemoglobin HBA

Availability Cobas b101 HbA1c devices are available at both Taranaki Base and Hawera hospital.
Turnaround Time 6 minutes from placing cartridge into analyser with patients' blood sample
Stability – Add test 60 seconds once blood sample has been applied to cartridge disc.

EDTA samples within 2 hours from collection Lithium-Heparin within 8 hours from collection

Specimen Requirements

Specimen Whole Blood  
Preferred Finger Stab - Droplet of Whole blood
Acceptable Purple top (EDTA) or Lithium-Heparin, non-gel (green top)
Unacceptable Samples containing anticoagulants or other additives

Separated samples

Minimum Volume 2µL whole blood
Specimen Collection and Transportation Clean and dry patients' hand and wipe away the first drop of blood.

Test Information

Test Method Cobas b101 - ROCHE analyser
Unit of Measurement mmol/mol
Reference Intervals Non-diabetic reference interval 20-40 mmol/mol.

Risk of microvascular complications increases exponentially above 55 mmol/mol.

If >/= 50 mmol/mol and not known to have diabetes, a repeat level >/= 50 after an interval supports the diagnosis.

If result is >65 mmol/mol, consider additional action.

Uncertainty of measurement N/A
Additional Information Ranges of measurement 20-130mmol/mol

The presence of some haemoglobin variants may require testing via alternative methods for HbA1c or the laboratory may refer clinician to glucose based criteria for diagnosis and monitoring.

Comments

Quick user guides for testing should be available with or near the analysers.

All procedures, worksheets and competency checklist for sign-off are available on the intranet to all Health NZ Staff.

This is an IANZ accredited test.

Owner Pam Chung (Pchung)
Last reviewed 1 Oct 2025
Next review Overdue, 1 Oct 2026