Troponin I i-stat (POCT)
Tests:Troponin i i-stat (POCT)
Overview | |
|---|---|
| Department | Hawera Lab Point of Care Testing (POCT) |
| Test Code | POCT |
| Synonyms | Trop I
Tni I-stat, hs-Tni POCT |
| Availability | Monday - Sunday at Hawera Hospital, available also out of hours for emergencies |
| Turnaround Time | 30 minutes, 15 minutes for test to analyse |
| Stability – Add test | 4 hours |
Specimen Requirements | |
| Specimen | Whole blood sample collected into Heparinized Tube |
| Preferred | Lithium Heparin Tube (Green Top) |
| Acceptable | Heparinized Blood Gas syringe |
| Unacceptable | Underfilled tubes, specimens must be filled to capacity.
Grossly haemolysed samples are unacceptable. Any paediatric green bullets containing gel. Any sample with a clot is unacceptable. |
| Minimum Volume | Full Green Top tube or full to line Blood Gas Syringe |
| Specimen Collection and Transportation | |
Test Information | |
| Test Method | i-Stat - hs-TnI cartridge (cardiac troponin I) |
| Unit of Measurement | ug/L |
| Reference Intervals | Reportable Range: 2.9 - 999.9 ng/L
Out of range: L2.9 or G1000 |
| Uncertainty of measurement | |
| Additional Information | The result is to be interpreted by the clinician in conjunction with the ED ACS Pathway |
Comments | |
| Quick user guides for testing should be available with the meter.
All procedures, worksheets and competency checklist for sign-off are available on the intranet to all Health NZ Staff. This test is covered under the current IANZ Laboratory accreditation. | |
| Owner | Pam Chung (Pchung) |
|---|---|
| Last reviewed | 10 Oct 2026 |
| Next review | 10 Oct 2027 OK |