Tests:Troponin i i-stat (POCT): Difference between revisions

From Taranaki Lab Manual
Tests:Troponin i i-stat (POCT)
Jump to navigation Jump to search
Pam Chung (talk | contribs)
No edit summary
Pam Chung (talk | contribs)
No edit summary
Line 25: Line 25:
|-
|-
|'''Stability – Add test'''
|'''Stability – Add test'''
|N/A
|30 minutes
|-
|-
! colspan="2" |
! colspan="2" |
Line 34: Line 34:
|-
|-
|'''Preferred'''
|'''Preferred'''
|Heparinized whole blood or plasma samples collected in plastic syringes or evacuated tubes containing lithium or sodium heparin filled to capacity.
|Lithium Heparin Tube (Green Top)
|-
|-
|'''Acceptable'''
|'''Acceptable'''
|Heparinized whole blood or plasma samples collected in plastic syringes or evacuated tubes containing lithium or sodium heparin filled to capacity.
|Heparinized Blood Gas syringe
|-
|-
|'''Unacceptable'''
|'''Unacceptable'''
|Grossly haemolysed samples are unacceptable
|Grossly haemolysed samples are unacceptable.
Any paediatric green bullets containing gel with a clot
Any Blood Gas syringe samples containing a clot
Any sample with a clot is unacceptable
|-
|-
|'''Minimum Volume'''
|'''Minimum Volume'''
|17 μL whole blood
|3ml Green Top
|-
|-
|'''Specimen Collection and Transportation'''
|'''Specimen Collection and Transportation'''

Revision as of 14:51, 24 December 2025

Overview

Department Point of Care Testing (POCT)
Test Code N/A
Synonyms Trop I

Tni I-stat, cTni

Availability Monday - Sunday at Hawera Hospital, available also out of hours for emergencies
Turnaround Time 10 minutes once cartridge is loaded into the i-Stat
Stability – Add test 30 minutes

Specimen Requirements

Specimen Venous sample collected into Heparinized
Preferred Lithium Heparin Tube (Green Top)
Acceptable Heparinized Blood Gas syringe
Unacceptable Grossly haemolysed samples are unacceptable.

Any paediatric green bullets containing gel with a clot Any Blood Gas syringe samples containing a clot Any sample with a clot is unacceptable

Minimum Volume 3ml Green Top
Specimen Collection and Transportation

Test Information

Test Method i-Stat - cTnI cartridge (cardiac troponin I)
Unit of Measurement ng/mL
Reference Intervals Reportable range 0.00-50.00ng/mL

<0.03 ng/mL

>0.03 ng/mL                 

Uncertainty of measurement N/A
Additional Information For information on the interpretation of POCT Troponin I results refer to the RACPP (Rural Accelerated Chest Pain Pathway)   

Rapid triage of chest pain patients, including those with acute coronary syndromes (ACS)

Comments

Quick user guides for testing should be available with the meter.

All procedures, worksheets and competency checklist for sign-off are available on the intranet to all Health NZ Staff.

This test is covered under the current IANZ Laboratory accreditation.

Owner Pam Chung (Pchung)
Last reviewed 24.12.2025
Next review 24.12.2026 (OK)