Troponin I i-stat (POCT)

From Taranaki Lab Manual
Tests:Troponin i i-stat (POCT)
Jump to navigation Jump to search

Overview

Department Point of Care Testing (POCT)
Test Code N/A
Synonyms Trop I

Tni I-stat, cTni

Availability Monday - Sunday at Hawera Hospital, available also out of hours for emergencies
Turnaround Time 10 minutes once cartridge is loaded into the i-Stat
Stability – Add test 30 minutes

Specimen Requirements

Specimen Whole blood sample collected into Heparinized Tube
Preferred Lithium Heparin Tube (Green Top)
Acceptable Heparinized Blood Gas syringe
Unacceptable Grossly haemolysed samples are unacceptable.

Any paediatric green bullets containing gel without a clot.

Any sample with a clot is unacceptable

Minimum Volume 3ml Green Top or 1ml Blood Gas Syringe
Specimen Collection and Transportation

Test Information

Test Method i-Stat - cTnI cartridge (cardiac troponin I)
Unit of Measurement ug/L
Reference Intervals Reportable range 0.00-50.00ng/mL

<0.03 ug/L

>0.03 ug/L                 

Uncertainty of measurement N/A
Additional Information For information on the interpretation of POCT Troponin I results refer to the RACPP (Rural Accelerated Chest Pain Pathway)   

Rapid triage of chest pain patients, including those with acute coronary syndromes (ACS)

Comments

Quick user guides for testing should be available with the meter.

All procedures, worksheets and competency checklist for sign-off are available on the intranet to all Health NZ Staff.

This test is covered under the current IANZ Laboratory accreditation.

Owner Pam Chung (Pchung)
Last reviewed 30.03.2026
Next review 30.03.2027 (OK)