| Line 61: | Line 61: | ||
|- | |- | ||
|'''Reference Intervals''' | |'''Reference Intervals''' | ||
| | |National diagnostic threshold for HbA1c (as of July 1st 2026) | ||
- '''Normal:''' <42 mmol/mol | |||
- '''Prediabetes:''' 42-47 mmol/mol | |||
- '''Diabetes''': ≥48 mmol/mol | |||
|- | |- | ||
|'''Uncertainty of measurement''' | |'''Uncertainty of measurement''' | ||
| Line 72: | Line 72: | ||
|- | |- | ||
|'''Additional Information''' | |'''Additional Information''' | ||
|Ranges of measurement 20- | |No confirmatory test now required if ≥53 mmol/mol | ||
Confirmatory test required as soon as practical if HbA1c 48-52 mmol/mol eg repeat HbA1c, fasting glucose or random glucose (if symptomatic) | |||
Ranges of measurement 20-130 mmol/mol | |||
The presence of some haemoglobin variants may require testing via alternative methods for HbA1c or the laboratory may refer clinician to glucose based criteria for diagnosis and monitoring. | The presence of some haemoglobin variants may require testing via alternative methods for HbA1c or the laboratory may refer clinician to glucose based criteria for diagnosis and monitoring. | ||
|- | |- | ||
| Line 87: | Line 91: | ||
| review_ping = | | review_ping = | ||
}} | }} | ||
<!-- review-ping: 2026-09-01T21:13:58.707Z --> | |||
<!-- review-ping: 2026-09-01T21:16:20.808Z --> | |||
Latest revision as of 10:16, 2 September 2026
Overview | |
|---|---|
| Department | Point of Care Testing (POCT) |
| Test Code | N/A |
| Synonyms | HbA1c whole blood (POCT)
glycated haemoglobin HBA |
| Availability | Cobas b101 HbA1c devices are available at both Taranaki Base and Hawera hospital. |
| Turnaround Time | 6 minutes from placing cartridge into analyser with patients' blood sample |
| Stability – Add test | 60 seconds once blood sample has been applied to cartridge disc.
EDTA samples within 2 hours from collection Lithium-Heparin within 8 hours from collection |
Specimen Requirements | |
| Specimen | Whole Blood |
| Preferred | Finger Stab - Droplet of Whole blood |
| Acceptable | Purple top (EDTA) or Lithium-Heparin, non-gel (green top) |
| Unacceptable | Samples containing anticoagulants or other additives
Separated samples |
| Minimum Volume | 2µL whole blood |
| Specimen Collection and Transportation | Clean and dry patients' hand and wipe away the first drop of blood. |
Test Information | |
| Test Method | Cobas b101 - ROCHE analyser |
| Unit of Measurement | mmol/mol |
| Reference Intervals | National diagnostic threshold for HbA1c (as of July 1st 2026)
- Normal: <42 mmol/mol - Prediabetes: 42-47 mmol/mol - Diabetes: ≥48 mmol/mol |
| Uncertainty of measurement | N/A |
| Additional Information | No confirmatory test now required if ≥53 mmol/mol
Confirmatory test required as soon as practical if HbA1c 48-52 mmol/mol eg repeat HbA1c, fasting glucose or random glucose (if symptomatic) Ranges of measurement 20-130 mmol/mol The presence of some haemoglobin variants may require testing via alternative methods for HbA1c or the laboratory may refer clinician to glucose based criteria for diagnosis and monitoring. |
Comments | |
| Quick user guides for testing should be available with or near the analysers.
All procedures, worksheets and competency checklist for sign-off are available on the intranet to all Health NZ Staff. This test is covered under the current IANZ Laboratory accreditation. | |
| Owner | Pam Chung (Pchung) |
|---|---|
| Last reviewed | 2 Sep 2026 |
| Next review | 2 Sep 2027 OK |