Tests:Blood gas Analysis (POCT): Difference between revisions
- Department
- Point of care
- Also at
- Hawera
Tests:Blood gas Analysis (POCT)
Created page with "{{DISPLAYTITLE: Blood Gas Analysis (POCT)}} Full Test List Point of Care Tests" |
m Mark as reviewed |
||
| (8 intermediate revisions by 3 users not shown) | |||
| Line 2: | Line 2: | ||
[[Category:Tests]] | [[Category:Tests]] | ||
[[Category:POCT]] | [[Category:POCT]] | ||
[[Category:Hawera]] | |||
{| class="wikitable" | |||
! colspan="2" | | |||
== '''Overview''' == | |||
|- | |||
|'''Department''' | |||
|Point of Care (POCT) | |||
|- | |||
|'''Test Code''' | |||
|N/A | |||
|- | |||
|'''Synonyms''' | |||
|BGAS, ABG, VBG | |||
|- | |||
|'''Availability''' | |||
|The analysers are available at both Taranaki Base and Hawera Hospital at various locations - Base ED, Theatres, ICU, NNU and Hawera ED | |||
|- | |||
|'''Turnaround Time''' | |||
|Results within 1 minute from sampling on the analyser | |||
|- | |||
|'''Stability – Add test''' | |||
|Sample must be processed within 30 minutes of collection. | |||
|- | |||
! colspan="2" | | |||
== '''Specimen Requirements''' == | |||
|- | |||
|'''Specimen''' | |||
|whole blood | |||
|- | |||
|'''Preferred''' | |||
|1 mL in a Radiometer safePICO syringe or a full 45μL Radiometer safeCLINITUBES Capillary | |||
|- | |||
|'''Acceptable''' | |||
|0.5 mL in a heparinised syringe | |||
|- | |||
|'''Unacceptable''' | |||
|Clotted samples | |||
Haemolysed blood samples | |||
Samples with large air gaps | |||
|- | |||
|'''Minimum Volume''' | |||
|0.2 mL whole blood in a 1 mL syringe | |||
|- | |||
|'''Specimen Transportation''' | |||
|ASAP from patient collection to analyser. Delays can compromise sample integrity and provide inaccurate results | |||
|- | |||
|'''Paediatric Requirements''' | |||
|A full 45μL collected in a Radiometer safeCLINITUBES Capillary (short) or a full 100μL whole blood collect in Laboratory provided capillary tube (long). | |||
0.2 mL whole blood in a 1 mL syringe | |||
|- | |||
! colspan="2" | | |||
== '''Test Information''' == | |||
|- | |||
|'''Test Method''' | |||
|Radiometer ABL90 FLEX PLUS blood gas analyser | |||
|- | |||
|'''Units of Measurement''' | |||
|Refer to the Laboratory | |||
|- | |||
|'''Reference Intervals''' | |||
|Refer to the Laboratory | |||
|- | |||
|'''Uncertainty of Measurement''' | |||
|Refer to the Laboratory | |||
|- | |||
|'''Additional Information''' | |||
|Blood gas panel includes: pH, pCO2, pO2, HCO3, Base Excess, Hb, sO2, Na, K, ICa, Cl, Anion Gap, Glucose, Lactate, Carboxyhaemoglobin and Methaemoglobin. | |||
|- | |||
! colspan="2" | | |||
== Comments == | |||
|- | |||
| colspan="2" |Quick user guides for running samples are available near the analysers. | |||
All procedures and competency checklist for sign-off are available on the intranet for all Health NZ staff. | |||
This test is covered under the current IANZ Laboratory accreditation. | |||
|} | |||
{{DocMeta | |||
| owner =Pchung | |||
| review_ping = | |||
}} | |||
<!-- review-ping: 2026-09-01T20:48:57.887Z --> | |||
Latest revision as of 09:48, 2 September 2026
Overview | |
|---|---|
| Department | Point of Care (POCT) |
| Test Code | N/A |
| Synonyms | BGAS, ABG, VBG |
| Availability | The analysers are available at both Taranaki Base and Hawera Hospital at various locations - Base ED, Theatres, ICU, NNU and Hawera ED |
| Turnaround Time | Results within 1 minute from sampling on the analyser |
| Stability – Add test | Sample must be processed within 30 minutes of collection. |
Specimen Requirements | |
| Specimen | whole blood |
| Preferred | 1 mL in a Radiometer safePICO syringe or a full 45μL Radiometer safeCLINITUBES Capillary |
| Acceptable | 0.5 mL in a heparinised syringe |
| Unacceptable | Clotted samples
Haemolysed blood samples Samples with large air gaps |
| Minimum Volume | 0.2 mL whole blood in a 1 mL syringe |
| Specimen Transportation | ASAP from patient collection to analyser. Delays can compromise sample integrity and provide inaccurate results |
| Paediatric Requirements | A full 45μL collected in a Radiometer safeCLINITUBES Capillary (short) or a full 100μL whole blood collect in Laboratory provided capillary tube (long).
0.2 mL whole blood in a 1 mL syringe |
Test Information | |
| Test Method | Radiometer ABL90 FLEX PLUS blood gas analyser |
| Units of Measurement | Refer to the Laboratory |
| Reference Intervals | Refer to the Laboratory |
| Uncertainty of Measurement | Refer to the Laboratory |
| Additional Information | Blood gas panel includes: pH, pCO2, pO2, HCO3, Base Excess, Hb, sO2, Na, K, ICa, Cl, Anion Gap, Glucose, Lactate, Carboxyhaemoglobin and Methaemoglobin. |
Comments | |
| Quick user guides for running samples are available near the analysers.
All procedures and competency checklist for sign-off are available on the intranet for all Health NZ staff. This test is covered under the current IANZ Laboratory accreditation. | |
| Owner | Pam Chung (Pchung) |
|---|---|
| Last reviewed | 2 Sep 2026 |
| Next review | 2 Sep 2027 OK |