Tests:Urinalysis (POCT): Difference between revisions
- Department
- Microbiology
- Also at
- Hawera
Tests:Urinalysis (POCT)
No edit summary |
m Mark as reviewed |
||
| (4 intermediate revisions by 3 users not shown) | |||
| Line 2: | Line 2: | ||
[[Category:Tests]] | [[Category:Tests]] | ||
[[Category:POCT]] | [[Category:POCT]] | ||
[[Category:Microbiology] | [[Category:Microbiology]] | ||
[[Category:Hawera]] | |||
{| class="wikitable" | {| class="wikitable" | ||
! colspan="2" | | ! colspan="2" | | ||
| Line 51: | Line 53: | ||
|'''Specimen Collection and Transportation''' | |'''Specimen Collection and Transportation''' | ||
|Urine collected in a clean dry pottle/container. | |Urine collected in a clean dry pottle/container. | ||
if being tested straight away sample can be ambient (8-24°C), If longer, then sample should be stored chilled (2-8°C) | if being tested straight away sample can be ambient (8-24°C), If longer, then sample should be stored chilled (2-8°C). Ideally tested within 4 hrs according to RCPA | ||
|- | |- | ||
! colspan="2" | | ! colspan="2" | | ||
| Line 60: | Line 62: | ||
|- | |- | ||
|'''Unit of Measurement''' | |'''Unit of Measurement''' | ||
|Varies depending on the analytes (mg/L, pH). Contact the Microbiology Department if more information is required | |Varies depending on the analytes (mg/L, pH). | ||
Results are reported in most cases as semi- quantitative. | |||
Contact the Laboratory - Microbiology Department if more information is required | |||
|- | |- | ||
|'''Reference Intervals''' | |'''Reference Intervals''' | ||
| | |N/A | ||
|- | |- | ||
|'''Uncertainty of measurement''' | |'''Uncertainty of measurement''' | ||
| Line 69: | Line 73: | ||
|- | |- | ||
|'''Additional Information''' | |'''Additional Information''' | ||
| | | | ||
|- | |- | ||
! colspan="2" | | ! colspan="2" | | ||
== '''Comments''' == | == '''Comments''' == | ||
|- | |- | ||
| colspan="2" |Quick user guides for testing should be available near the | | colspan="2" |Quick user guides for testing should be available near the analysers | ||
All procedures, worksheets and competency checklist for sign-off are available on the intranet to all Health NZ Staff. | All procedures, worksheets and competency checklist for sign-off are available on the intranet to all Health NZ Staff. | ||
| Line 87: | Line 88: | ||
| review_ping = | | review_ping = | ||
}} | }} | ||
<!-- review-ping: 2026-09-01T20:42:35.130Z --> | |||
<!-- review-ping: 2026-09-01T21:19:23.978Z --> | |||
Latest revision as of 10:19, 2 September 2026
Overview | ||
|---|---|---|
| Department | Point of Care Testing (POCT) | |
| Test Code | N/A | |
| Synonyms | Dipstick, urine
Urinalysis Glucose, urine dipstick Ketones, urine dipstick pH, urine dipstick Protein, urine dipstick | |
| Availability | Urine dipstick readers (Urisys 1100) are available at both Taranaki Base and Hawera ED | |
| Turnaround Time | 1 minutes from the time the urine dipstick is placed on the device | |
| Stability – Add test | ASAP if at room temperature | |
Specimen Requirements | ||
| Specimen | Urine | |
| Preferred | Fresh urine collected in a clean dry pottle/container. Ideally a Midstream urine collected anytime. This is to limit any external contamination. | |
| Acceptable | Specimens that have been stored chilled in the refrigerator (2-8°C) | |
| Minimum Volume | 2ml | |
| Specimen Collection and Transportation | Urine collected in a clean dry pottle/container.
if being tested straight away sample can be ambient (8-24°C), If longer, then sample should be stored chilled (2-8°C). Ideally tested within 4 hrs according to RCPA | |
Test Information | ||
| Test Method | Roche Urisys 1100 analyser and in conjunction with Roche Combur 10 Test UX Dipsticks | |
| Unit of Measurement | Varies depending on the analytes (mg/L, pH).
Results are reported in most cases as semi- quantitative. Contact the Laboratory - Microbiology Department if more information is required | |
| Reference Intervals | N/A | |
| Uncertainty of measurement | N/A | |
| Additional Information | ||
Comments | ||
| Quick user guides for testing should be available near the analysers
All procedures, worksheets and competency checklist for sign-off are available on the intranet to all Health NZ Staff. This test is covered under the current IANZ Laboratory accreditation. |
||
| Owner | Pam Chung (Pchung) |
|---|---|
| Last reviewed | 2 Sep 2026 |
| Next review | 2 Sep 2027 OK |