|Test Availability
|Base Lab Availability
|Monday - Sunday, testing routinely available out of hours  
|Monday - Sunday, testing routinely available out of hours  
|-
|Hawera Lab Availability
|Monday - Friday 8 am - 5pm
|-
|-
|Turnaround Time
|Turnaround Time
Line 30: Line 35:
== Specimen Requirements ==
== Specimen Requirements ==
|-
|-
|Specimen Type
|Specimen  
|Plasma
|Plasma
|-
|-
Line 54: Line 59:
== Aliquot Requirements ==
== Aliquot Requirements ==
|-
|-
|Sample Type
|Sample  
|
|
|-
|-
Line 84: Line 89:
== Comments ==
== Comments ==
|-
|-
| colspan="2" |Included in coagulation screen.  
| colspan="2" | Cryoprecipitate transfusion is indicated if fibrinogen is <1.0 g/L in the presence of clinical bleeding.  
 
Cryoprecipitate transfusion is indicated if fibrinogen is <1.0 g/L in the presence of clinical bleeding.  
|}
|}
{{DocMeta
| owner = Mabbett1
| review_ping =
}}

Latest revision as of 13:11, 12 September 2025

Included in coagulation screen

Overview

Department Haematology    
Test Code FIB
Synonyms Coag screen

DIC screen  

Base Lab Availability Monday - Sunday, testing routinely available out of hours
Hawera Lab Availability Monday - Friday 8 am - 5pm
Turnaround Time 1 – 4 hrs    
Stability – Add test   24 hrs

Specimen Requirements

Specimen Plasma
Preferred 2.7 mL na citrate Blue-Top
Acceptable
Unacceptable Clotted samples

Under or overfilled tubes Haemolysed Samples older than 24 hours

Specimen Transportation Ambient  
Paediatric Requirements   1.0 mL Blue-Top    

Aliquot Requirements

Sample
Preferred Volume
Minimum Volume
Aliquot Transportation

Clinical Information

Reference Intervals Reference Range  1.5 - 4.0 g/L  
Test Method Von Clauss method using mechanical clot detection, performed on Stago Compact Max2 Analyser  
Advanced Notice Not required    
Patient Preparation Not required    

Comments

 Cryoprecipitate transfusion is indicated if fibrinogen is <1.0 g/L in the presence of clinical bleeding.
Owner Craig Mabbett (Mabbett1)
Last reviewed 12 Sep 2025