Jump to content
Health NZ Taranaki Lab Manual

Tests:HbA1c (POCT): Difference between revisions

Department
Point of care
Tests:HbA1c (POCT)
Pam Chung (talk | contribs)
No edit summary
Pam Chung (talk | contribs)
No edit summary
Line 14: Line 14:
|'''Synonyms'''
|'''Synonyms'''
|HbA1c whole blood (POCT)
|HbA1c whole blood (POCT)
glycated haemoglobin
HBA
|-
|-
|'''Availability'''
|'''Availability'''
|Cobas b101 HbA1c devices are available at both Taranaki Base and Hawera hospital.
|Cobas b101 HbA1c devices are available at both Taranaki Base and Hawera hospital.
The Ketone meters <u>must</u> be returned to the Laboratory after use for others to use and QC.
|-
|-
|'''Turnaround Time'''
|'''Turnaround Time'''
|As and when the Ketone meters are available.
|6 minutes from placing cartridge into analyser with patients' blood sample
Results once tested - 10 seconds
|-
|-
|'''Stability – Add test'''
|'''Stability – Add test'''
|N/A
|60 seconds once blood sample has been applied to cartridge disc.
EDTA samples within 2 hours from collection
Lithium-Heparin within 8 hours from collection
|-
|-
! colspan="2" |
! colspan="2" |
Line 36: Line 38:
|-
|-
|'''Acceptable'''
|'''Acceptable'''
|Blood Gas, Purple top (EDTA), 0.6 µL capillary sample collected directly onto a Statstrip test strip
|Purple top (EDTA) or Lithium-Heparin, non-gel (green top)
|-
|-
|'''Unacceptable'''
|'''Unacceptable'''
|Samples older than 4 hours
|Samples containing anticoagulants or other additives
Neonate samples
 
Separated samples
Separated samples
|-
|-
|'''Minimum Volume'''
|'''Minimum Volume'''
|0.6 µL whole blood
|2µL whole blood
|-
|-
|'''Specimen Collection and Transportation'''
|'''Specimen Collection and Transportation'''
Line 54: Line 54:
|-
|-
|'''Test Method'''
|'''Test Method'''
|Nova StatStrip Xpress 1/2 handheld device and Ketone StatStrip
|Cobas b101 - ROCHE analyser
|-
|-
|'''Unit of Measurement'''
|'''Unit of Measurement'''
| mmol/L
| mmol/mol
|-
|-
|'''Reference Intervals'''
|'''Reference Intervals'''
|<0.3 mmol/L
|Non-diabetic reference interval 20-40 mmol/mol.
Risk of microvascular complications increases exponentially above 55 mmol/mol.
 
If >/= 50 mmol/mol and not known to have diabetes, a repeat level >/= 50 after an interval supports the diagnosis.
 
If result is >65 mmol/mol, consider additional action.
|-
|-
|'''Uncertainty of measurement'''
|'''Uncertainty of measurement'''
Line 66: Line 71:
|-
|-
|'''Additional Information'''
|'''Additional Information'''
|Ketones include acetoacetic acid, beta-hydroxybutyric acid and acetone. The relative proportions of all ketones depend on the severity of the ketosis and the availability of NADH for metabolism. Beta-hydroxybutyric acid (OHB) can be added up to 4 hrs pos.
|Ranges of measurement 20-130mmol/mol
Ketone Measuring range 0.1 -7.0mmol/L
The presence of some haemoglobin variants may require testing via alternative methods for HbA1c or the laboratory may refer clinician to glucose based criteria for diagnosis and monitoring.
|-
|-
! colspan="2" |
! colspan="2" |
== '''Comments''' ==
== '''Comments''' ==
|-
|-
| colspan="2" |Quick user guides for testing should be available with the meter.
| colspan="2" |Quick user guides for testing should be available with or near the analysers.
All procedures, worksheets and competency checklist for sig off are available to all Health NZ Staff.
All procedures, worksheets and competency checklist for sign-off are available on the intranet to all Health NZ Staff.
|}
|}
{{DocMeta
{{DocMeta

Revision as of 14:17, 1 October 2025

Overview

Department POCT
Test Code N/A
Synonyms HbA1c whole blood (POCT)

glycated haemoglobin HBA

Availability Cobas b101 HbA1c devices are available at both Taranaki Base and Hawera hospital.
Turnaround Time 6 minutes from placing cartridge into analyser with patients' blood sample
Stability – Add test 60 seconds once blood sample has been applied to cartridge disc.

EDTA samples within 2 hours from collection Lithium-Heparin within 8 hours from collection

Specimen Requirements

Specimen Whole Blood  
Preferred Finger Stab - Droplet of Whole blood
Acceptable Purple top (EDTA) or Lithium-Heparin, non-gel (green top)
Unacceptable Samples containing anticoagulants or other additives

Separated samples

Minimum Volume 2µL whole blood
Specimen Collection and Transportation Clean and dry patients' hand and wipe away the first drop of blood.

Test Information

Test Method Cobas b101 - ROCHE analyser
Unit of Measurement mmol/mol
Reference Intervals Non-diabetic reference interval 20-40 mmol/mol.

Risk of microvascular complications increases exponentially above 55 mmol/mol.

If >/= 50 mmol/mol and not known to have diabetes, a repeat level >/= 50 after an interval supports the diagnosis.

If result is >65 mmol/mol, consider additional action.

Uncertainty of measurement N/A
Additional Information Ranges of measurement 20-130mmol/mol

The presence of some haemoglobin variants may require testing via alternative methods for HbA1c or the laboratory may refer clinician to glucose based criteria for diagnosis and monitoring.

Comments

Quick user guides for testing should be available with or near the analysers.

All procedures, worksheets and competency checklist for sign-off are available on the intranet to all Health NZ Staff.

Owner PChung
Last reviewed 1 Oct 2025
Next review Overdue, 1 Oct 2026