Tests:HbA1c (POCT): Difference between revisions
- Department
- Point of care
- Also at
- Hawera
No edit summary |
No edit summary |
||
| Line 2: | Line 2: | ||
[[Category:Tests]] | [[Category:Tests]] | ||
[[Category:POCT]] | [[Category:POCT]] | ||
[[Category:Hawera]] | |||
{| class="wikitable" | {| class="wikitable" | ||
! colspan="2" | | ! colspan="2" | | ||
Revision as of 09:02, 3 October 2025
Overview | |
|---|---|
| Department | Point of Care Testing (POCT) |
| Test Code | N/A |
| Synonyms | HbA1c whole blood (POCT)
glycated haemoglobin HBA |
| Availability | Cobas b101 HbA1c devices are available at both Taranaki Base and Hawera hospital. |
| Turnaround Time | 6 minutes from placing cartridge into analyser with patients' blood sample |
| Stability – Add test | 60 seconds once blood sample has been applied to cartridge disc.
EDTA samples within 2 hours from collection Lithium-Heparin within 8 hours from collection |
Specimen Requirements | |
| Specimen | Whole Blood |
| Preferred | Finger Stab - Droplet of Whole blood |
| Acceptable | Purple top (EDTA) or Lithium-Heparin, non-gel (green top) |
| Unacceptable | Samples containing anticoagulants or other additives
Separated samples |
| Minimum Volume | 2µL whole blood |
| Specimen Collection and Transportation | Clean and dry patients' hand and wipe away the first drop of blood. |
Test Information | |
| Test Method | Cobas b101 - ROCHE analyser |
| Unit of Measurement | mmol/mol |
| Reference Intervals | Non-diabetic reference interval 20-40 mmol/mol.
Risk of microvascular complications increases exponentially above 55 mmol/mol. If >/= 50 mmol/mol and not known to have diabetes, a repeat level >/= 50 after an interval supports the diagnosis. If result is >65 mmol/mol, consider additional action. |
| Uncertainty of measurement | N/A |
| Additional Information | Ranges of measurement 20-130mmol/mol
The presence of some haemoglobin variants may require testing via alternative methods for HbA1c or the laboratory may refer clinician to glucose based criteria for diagnosis and monitoring. |
Comments | |
| Quick user guides for testing should be available with or near the analysers.
All procedures, worksheets and competency checklist for sign-off are available on the intranet to all Health NZ Staff. This test is covered under the current IANZ Laboratory accreditation. | |
| Owner | Pam Chung (Pchung) |
|---|---|
| Last reviewed | 3 Oct 2025 |
| Next review | Overdue, 3 Oct 2026 |