Jump to content
Health NZ Taranaki Lab Manual

Tests:Blood gas Analysis (POCT): Difference between revisions

Department
Point of care
Tests:Blood gas Analysis (POCT)
Pam Chung (talk | contribs)
No edit summary
Pam Chung (talk | contribs)
No edit summary
Line 7: Line 7:
|-
|-
|'''Department'''
|'''Department'''
|Biochemistry 
|Point of Care (POCT)
|-
|-
|'''Test Code'''
|'''Test Code'''
Line 14: Line 14:
|'''Synonyms'''
|'''Synonyms'''
|BGAS, ABG, VBG 
|BGAS, ABG, VBG 
|-
|'''Availability'''
|The analysers are available at both Taranaki Base and Hawera Hospital at various locations - Base ED, Theatres, ICU, NNU and Hawera ED
|-
|-
|'''Turnaround Time'''
|'''Turnaround Time'''
|
|Results within 1 minute from sampling on the analyser
|-
|-
|'''Stability – Add test'''
|'''Stability – Add test'''
|
|Sample must be processed within 30 minutes of collection. 
|-
|-
! colspan="2" |
! colspan="2" |
Line 28: Line 31:
|-
|-
|'''Preferred'''
|'''Preferred'''
|0.5 mL in heparinised syringe (see Blood Gas Syringe Collection Instructions) 
|1 mL in a Radiometer safePICO syringe or a full 45μL Radiometer safeCLINITUBES Capillary
|-
|-
|'''Acceptable'''
|'''Acceptable'''
|
|0.5 mL in a heparinised syringe
|-
|-
|'''Unacceptable'''
|'''Unacceptable'''
|Clotted samples
|Clotted samples
Samples with Needle attached
Samples with with large air gaps
Samples with large air gaps
|-
|'''Minimum Volume'''
|0.2 mL whole  blood in a 1 mL syringe  
|-
|-
|'''Specimen Transportation'''
|'''Specimen Transportation'''
|ASAP to lab – Sample must be processed within 30 minutes of collection. 
|ASAP from patient collection to analyser. Delays can compromise sample integrity and provide inaccurate results 
 
If delay anticipated samples can be stored on ice for up to 1hr, samples on ice are unacceptable for Potassium analysis. 
|-
|-
|'''Paediatric Requirements'''
|'''Paediatric Requirements'''
|0.1 mL whole blood in capillary tube.  (Please contact ext. 7881 for capillary tubes.)
|A full 45μL collected in a Radiometer safeCLINITUBES Capillary (short) or a full 100μL whole blood collect in Laboratory provided capillary tube (long).   
0.2 mL whole blood in a 1 mL syringe
0.2 mL whole blood in a 1 mL syringe
|-
|-
! colspan="2" |
! colspan="2" |
== Aliquot Requirements ==
== '''Test Information''' ==
|-
|-
|Sample
|'''Test Method'''
|0.5 mL whole blood  
|Radiometer ABL90 FLEX PLUS blood gas analyser 
|-
|-
|Preferred Volume
|'''Units of Measurement'''
|
|
|-
|-
|Minimum Volume
|'''Reference Intervals'''
|0.2 mL whole  blood in a 1 mL syringe  
|
|-
|Aliquot Transportation
|Sample must be kept anaerobic.  Samples with small air bubbles will be  processed with a comment.  Samples with  large air gaps in the syringe will not be processed. 
|-
! colspan="2" |
== Clinical Information ==
|-
|-
|Reference Intervals
|'''Uncertainty of Measurement'''
|
|
|-
|-
|Test Method
|'''Additional Information'''
|Radiometer ABL90 FLEX blood gas analyser 
|Blood gas panel includes:  pH, pCO2, pO2, HCO3, Base Excess, Hb, sO2,  Na, K, ICa, Cl, Anion Gap, Glucose, Lactate, Carboxyhaemoglobin and Methaemoglobin.
|-
|Advanced Notice
|Contact Laboratory  
|-
|Patient Preparation
|
|-
|-
! colspan="2" |
! colspan="2" |
== Comments ==
== Comments ==
|-
|-
| colspan="2" |Blood gas panel includes:  pH, pCO2, pO2, HCO3, Base Excess, Hb, sO2,  Na, K, ICa, Cl, Anion Gap, Glucose and Lactate.  
| colspan="2" |Quick user guides for running samples are available near the analysers.
 
All procedures and competency checklist for sign-off are available on the intranet for all Health NZ staff.
Carboxyhaemoglobin and Methaemoglobin are available on request.
This test is covered under the current IANZ Laboratory accreditation.
|}
|}
{{DocMeta
{{DocMeta

Revision as of 17:04, 1 October 2025

Overview

Department Point of Care (POCT)
Test Code N/A
Synonyms BGAS, ABG, VBG 
Availability The analysers are available at both Taranaki Base and Hawera Hospital at various locations - Base ED, Theatres, ICU, NNU and Hawera ED
Turnaround Time Results within 1 minute from sampling on the analyser
Stability – Add test Sample must be processed within 30 minutes of collection. 

Specimen Requirements

Specimen whole blood
Preferred 1 mL in a Radiometer safePICO syringe or a full 45μL Radiometer safeCLINITUBES Capillary
Acceptable 0.5 mL in a heparinised syringe
Unacceptable Clotted samples

Samples with with large air gaps

Minimum Volume 0.2 mL whole blood in a 1 mL syringe  
Specimen Transportation ASAP from patient collection to analyser. Delays can compromise sample integrity and provide inaccurate results
Paediatric Requirements A full 45μL collected in a Radiometer safeCLINITUBES Capillary (short) or a full 100μL whole blood collect in Laboratory provided capillary tube (long). 

0.2 mL whole blood in a 1 mL syringe

Test Information

Test Method Radiometer ABL90 FLEX PLUS blood gas analyser 
Units of Measurement
Reference Intervals
Uncertainty of Measurement
Additional Information Blood gas panel includes:  pH, pCO2, pO2, HCO3, Base Excess, Hb, sO2, Na, K, ICa, Cl, Anion Gap, Glucose, Lactate, Carboxyhaemoglobin and Methaemoglobin.

Comments

Quick user guides for running samples are available near the analysers.

All procedures and competency checklist for sign-off are available on the intranet for all Health NZ staff. This test is covered under the current IANZ Laboratory accreditation.

Owner Pam Chung (Pchung)
Last reviewed 1 Oct 2025
Next review Overdue, 1 Oct 2026