Jump to content
Health NZ Taranaki Lab Manual

Tests:Blood gas Analysis (POCT): Difference between revisions

Department
Point of care
Also at
Hawera
Tests:Blood gas Analysis (POCT)
No edit summary
m Added more details and deleted extra words
Line 40: Line 40:
|'''Unacceptable'''
|'''Unacceptable'''
|Clotted samples
|Clotted samples
Samples with with large air gaps
Haemolysed blood sample
Samples with large air gaps
|-
|-
|'''Minimum Volume'''
|'''Minimum Volume'''

Revision as of 09:46, 2 September 2026


Overview

Department Point of Care (POCT)
Test Code N/A
Synonyms BGAS, ABG, VBG 
Availability The analysers are available at both Taranaki Base and Hawera Hospital at various locations - Base ED, Theatres, ICU, NNU and Hawera ED
Turnaround Time Results within 1 minute from sampling on the analyser
Stability – Add test Sample must be processed within 30 minutes of collection. 

Specimen Requirements

Specimen whole blood
Preferred 1 mL in a Radiometer safePICO syringe or a full 45μL Radiometer safeCLINITUBES Capillary
Acceptable 0.5 mL in a heparinised syringe
Unacceptable Clotted samples

Haemolysed blood sample Samples with large air gaps

Minimum Volume 0.2 mL whole blood in a 1 mL syringe  
Specimen Transportation ASAP from patient collection to analyser. Delays can compromise sample integrity and provide inaccurate results
Paediatric Requirements A full 45μL collected in a Radiometer safeCLINITUBES Capillary (short) or a full 100μL whole blood collect in Laboratory provided capillary tube (long). 

0.2 mL whole blood in a 1 mL syringe

Test Information

Test Method Radiometer ABL90 FLEX PLUS blood gas analyser 
Units of Measurement Refer to the Laboratory
Reference Intervals Refer to the Laboratory
Uncertainty of Measurement Refer to the Laboratory
Additional Information Blood gas panel includes:  pH, pCO2, pO2, HCO3, Base Excess, Hb, sO2, Na, K, ICa, Cl, Anion Gap, Glucose, Lactate, Carboxyhaemoglobin and Methaemoglobin.

Comments

Quick user guides for running samples are available near the analysers.

All procedures and competency checklist for sign-off are available on the intranet for all Health NZ staff.

This test is covered under the current IANZ Laboratory accreditation.

Owner Pam Chung (Pchung)
Last reviewed 2 Sep 2026
Next review 2 Sep 2027 OK