Glucose (POCT)
Tests:Glucose (POCT)
Overview | |
|---|---|
| Department | POCT |
| Test Code | N/A |
| Synonyms | Glucose, blood |
| Availability | StatStrip handheld Glucose meters are available throughout Taranaki Base and Hawera Hospital. |
| Turnaround Time | 6 seconds |
| Stability – Add test | N/A |
Specimen Requirements | |
| Specimen | Whole Blood |
| Preferred | Finger Stab - Droplet of Whole blood |
| Acceptable | Blood Gas, Purple top (EDTA), 0.6 µL capillary sample collected directly onto a Statstrip test strip |
| Unacceptable | Samples older than 4 hours
Separated samples |
| Minimum Volume | 0.6 µL whole blood |
| Specimen Collection and Transportation | Clean and dry patients' hand and wipe away the first drop of blood. |
Test Information | |
| Test Method | Nova StatStrip Xpress 1/2 handheld device and Ketone StatStrip |
| Unit of Measurement | mmol/L |
| Reference Intervals | 3.5-5.5 mmol/L Fasting
3.5-7.7 mmol/L Non-Fasting |
| Uncertainty of measurement | N/A |
| Additional Information | Glucose Measuring range using the StatStrip meters 0.6-33.3mmol/L |
Comments | |
| Quick user guides for testing should be available with the meters
All procedures, worksheets and competency checklist for sig off are available to all Health NZ Staff This test is an IANZ accredited test | |
| Owner | Pam Chung (Pchung) |
|---|---|
| Last reviewed | 01.10.2025 |
| Next review | 01.10.2026 (OK) |