HbA1c (POCT)
- Department
- Point of care
Tests:HbA1c (POCT)
Overview | |
|---|---|
| Department | POCT |
| Test Code | N/A |
| Synonyms | HbA1c whole blood (POCT) |
| Availability | Cobas b101 HbA1c devices are available at both Taranaki Base and Hawera hospital.
The Ketone meters must be returned to the Laboratory after use for others to use and QC. |
| Turnaround Time | As and when the Ketone meters are available.
Results once tested - 10 seconds |
| Stability – Add test | N/A |
Specimen Requirements | |
| Specimen | Whole Blood |
| Preferred | Finger Stab - Droplet of Whole blood |
| Acceptable | Blood Gas, Purple top (EDTA), 0.6 µL capillary sample collected directly onto a Statstrip test strip |
| Unacceptable | Samples older than 4 hours
Neonate samples Separated samples |
| Minimum Volume | 0.6 µL whole blood |
| Specimen Collection and Transportation | Clean and dry patients' hand and wipe away the first drop of blood. |
Test Information | |
| Test Method | Nova StatStrip Xpress 1/2 handheld device and Ketone StatStrip |
| Unit of Measurement | mmol/L |
| Reference Intervals | <0.3 mmol/L |
| Uncertainty of measurement | N/A |
| Additional Information | Ketones include acetoacetic acid, beta-hydroxybutyric acid and acetone. The relative proportions of all ketones depend on the severity of the ketosis and the availability of NADH for metabolism. Beta-hydroxybutyric acid (OHB) can be added up to 4 hrs pos.
Ketone Measuring range 0.1 -7.0mmol/L |
Comments | |
| Quick user guides for testing should be available with the meter.
All procedures, worksheets and competency checklist for sig off are available to all Health NZ Staff. | |
| Owner | PChung |
|---|---|
| Last reviewed | 1 Oct 2025 |
| Next review | Overdue, 1 Oct 2026 |