Glucose (POCT)

From Taranaki Lab Manual
Revision as of 13:25, 1 October 2025 by Pam Chung (talk | contribs)

Tests:Glucose (POCT)
Jump to navigation Jump to search


Overview

Department POCT
Test Code N/A
Synonyms Glucose, blood 
Availability StatStrip handheld Glucose meters are available throughout Taranaki Base and Hawera Hospital.
Turnaround Time 6 seconds
Stability – Add test N/A

Specimen Requirements

Specimen Whole Blood  
Preferred Finger Stab - Droplet of Whole blood
Acceptable Blood Gas, Purple top (EDTA), 0.6 µL capillary sample collected directly onto a Statstrip test strip
Unacceptable Samples older than 4 hours

Separated samples

Minimum Volume 0.6 µL whole blood
Specimen Collection and Transportation Clean and dry patients' hand and wipe away the first drop of blood.

Test Information

Test Method Nova StatStrip Xpress 1/2 handheld device and Ketone StatStrip
Unit of Measurement mmol/L
Reference Intervals 3.5-5.5 mmol/L Fasting

3.5-7.7 mmol/L Non-Fasting

Uncertainty of measurement N/A
Additional Information Glucose Measuring range 0.6-33.3mmol/L

Comments

Quick user guides for testing should be available with the meters

All procedures, worksheets and competency checklist for sig off are available to all Health NZ Staff

Owner Pam Chung (Pchung)
Last reviewed 01.10.2025
Next review 01.10.2026 (OK)