D Dimer Assay, Plasma
- Department
- Haematology
Tests:D Dimer Assay, Plasma
Overview | |
|---|---|
| Department | Haematology |
| Test Code | DIME |
| Synonyms | XDP
FDP Part of DIC Screen |
| Availability | Monday - Sunday, testing routinely available out of hours |
| Turnaround Time | 2 hours |
| Stability – Add test | 24 hours |
Specimen Requirements | |
| Specimen | Plasma |
| Preferred | 2.7 mL Sodium Citrate Blue-Top |
| Acceptable | |
| Unacceptable | Clotted sample
Under or overfilled tubes Haemolysed Sample older than 24 hours |
| Specimen Transportation | |
| Paediatric Requirements | 1 mL Sodium Citrate Blue-Top |
Aliquot Requirements | |
| Sample | |
| Preferred Volume | |
| Minimum Volume | |
| Aliquot Transportation | |
Clinical Information | |
| Reference Intervals | Reference range: <500 µg/L |
| Test Method | Turbidometric immunoassay performed on Stago Compact Max2 Analyser |
| Advanced Notice | Not required |
| Patient Preparation | Not required |
Comments | |
| Clinical Reminder: The D-Dimer test is only useful for exclusion of VTE in patients with a low clinical probability of DVT or PE.
A negative result largely excludes VTE in this setting, while a positive result is non-specific | |