Anion Gap, Plasma

Tests:Anion Gap, Plasma


Overview

Department Biochemistry 
Test Code AGAP
Synonyms AGap

AG Calculation 

Test Availability Monday - Sunday, testing routinely available out of hours
Turnaround Time 1 – 2 hours 
Stability – Add test

Specimen Requirements

Specimen Plasma

Serum

Preferred 6 mL Plain Red-Top
Acceptable 6 mL Li-Hep Green-Top  
Unacceptable Haemolysed samples

Samples older than 4 hours.

Specimen Transportation
Paediatric Requirements 600 µL Microtainer tube  

Aliquot Requirements

Sample Plasma

Serum 

Preferred Volume
Minimum Volume 200 µL
Aliquot Transportation Keep anaerobic

Clinical Information

Reference Intervals Reference range: 10 - 20 nmol/L. 

Anion Gap Interpretive Comments:

For AG 21-60 mmol/l:

Increased anion gap; consider lactic acidosis, ketoacidosis, renal failure or ingestion of salicylate or methanol as possible causes. 

For AG 8-20 mmol/l AND HCO3- <20 mmol/l AND Cl- >107 mmol/l:

Normal anion gap acidosis: consider renal tubular acidosis, acetazolamide, toluene ingestion or gastrointestinal loss (diarrhoea).

Measurement of urinary pH and urinary anion gap will assist differential diagnosis. 

For K < 3.0 AND HCO3- >28 mmol/l AND AG 8-20 mmol/l:

Hypokalaemic alkalosis: consider renal loss (diuretics, Gitelmans or Bartters Syndrome) and gastrointestinal loss (vomiting, nasogastric suction).

Measurement of urinary electrolytes, urinary and serum calcium and magnesium will assist differential diagnosis. 

For AG <7 mmol/l:

Low anion gap - possible causes include hypoproteinaemia or paraprotein (consider protein electrophoresis). 

Test Method Calculation:(Na + K) - (Cl- + HCO3) = Anion Gap 
Advanced Notice Not required  
Patient Preparation Not required

Comments

Blood gas specimens routinely report Anion Gap.  
Owner Clair Crossley (Crossley1)
Last reviewed 12 Sep 2025
Next review Overdue, 12 Sep 2026